Recall: ISC UltraLink

Date of publication: November 30th 2023.

Following on from our inspection notice issued 17 November 2023, our team has been intensively investigating the potential risk and the root cause of the problem. Our investigation identified that the press fixture used to secure the assembly of the SafeLock button moved outside of specification, creating the potential for a faulty press fit of the button, which can lead to the button turning open and coming apart.

Stop use

Due to the risk of unintentional opening of the device, which could result in a life-threatening situation, we are issuing this notice instructing users to stop using their device, if it falls within the range of serial numbers:

LK100A1.2: 170431 - 177209; or LK101A1.2: 170433 - 176974.

STOP USING ANY OF THE ABOVE ITEMS IMMEDIATELY AND REMOVE THEM FROM SERVICE.

Manufacturing corrective actions

Manufacturing corrective actions have been introduced to repair and ensure that the press fixture remains within specification; and additional procedures have been introduced to ensure that the SafeLock buttons are secured correctly. Validation testing has been successfully completed and production of UltraLinks has resumed, with this new process in place.

Contact

DEALER RECALL: Please contact your Customer Service Representative to register your inventory stock for recall.

END USER RECALL: Share this link with your customers so that they can register their device(s) for recall: http://eepurl.com/iE_-8s.

Source

This information is provided and published by ISC. For more information of this safety warning, please visit: the webpage of ISC or contact us.